Imager Sensitivity Settings May Affect Detection of Chemiluminescent Western Blot Results


  • troubleshooting chemiluminescent western blots
  • troubleshooting western blot
  • imager sensitivity
  • Standard and High Sensitivity Settings on the C-DiGit Blot Scanner

    Making sure that the sensitivity setting is optimal to capture the most signal from your chemiluminescent Western blot could be the difference between getting a good, strong signal or getting a signal that you can barely see. This is our possible cause 7 for weak chemiluminescent signals.

    How can you avoid possible cause 7? On the C-DiGit® Blot Scanner, use High Sensitivity setting (12-min scan) for more sensitive detection. On the Odyssey® Fc Dual-Mode Imaging System, use a longer integration time (up to 10 min). 

    Why is this important? Well, digital imaging with the C-DiGit Blot Scanner or Odyssey Fc Imager will not generally reach a saturation point. Begin with a longer acquisition time to ensure best sensitivity, then optimize to shorter scan times. 

    In Table 1 below, we tested the performance differences of a Western blot detected with SuperSignal® West Dura on the C-DiGit Blot Scanner when the same blot is imaged on High Sensitivity (12 min scan) versus Standard Sensitivity (6 min scan). As you can see, the longer scan time and higher sensitivity make a big difference in the results.


    Table 1Optimal BlotSatisfactory Blot
    Images
    SubstrateSuperSignal West Dura1SuperSignal West Dura
    SensitivityHigh (12 min)Standard (6 min)
    PerformanceSignal – 12,300Signal – 5,030

    1Comparable to WesternSure® PREMIUM Chemiluminescent Substrate So be sure to check your sensitivity settings before you scan!

    Other posts related to troubleshooting chemiluminescent Western blots: